Comprehensive and Detailed Explanation From Exact Extract:
Mechanical Index (MI) estimates the likelihood of non-thermal bioeffects such as cavitation (formation of microbubbles).
Principles and Instrumentation state:
"Mechanical Index reflects the risk of cavitation. Higher MI values increase cavitation potential."
Attenuation coefficient (A) refers to energy loss.
Thermal index (C) reflects potential heating.
Frame rate (D) relates to temporal resolution.
Therefore, the correct answer is B: Mechanical index.
—
QUESTION NO42
Which target group is used to evaluate transverse distance measurement accuracy in this tissue-mimicking phantom image?
A ultrasound image of a baby
Description automatically generated with medium confidence
A. A
B. B
C. C
D. D
Answer: C
Comprehensive and Detailed Explanation From Exact Extract:
In a tissue-mimicking phantom, the group of targets aligned side-by-side (horizontally) is used to evaluate transverse (lateral) distance measurement accuracy. In the image, target group C (in red) represents the lateral spacing targets.
Principles and Instrumentation state:
"Transverse distance accuracy is evaluated using laterally spaced targets in the phantom to verify correct lateral measurement calibration."
Group A (top) is typically for near field assessment.
Group B may be for axial resolution or cyst evaluation.
Group D (deep) may assess depth calibration or sensitivity.
Therefore, the correct answer is C: C.
—